Australia – Time, Cost, and Complexity of Registration
| Device classification in Australia | Class I
Non-sterile; Non-measuring |
Class I
Sterile or Measuring |
Class IIa | Class IIb | Class III |
| How long you should expect to wait after submission until approval is granted | 2 to 5 business days | up to 10 business days | between 2 and 6 months | 3 to 9 months | 6 to 12 months |
| Validity period for device registrations/ARTG listing | The Australian Register of Therapeutic Goods (ARTG) listing does not have a formal expiry date. The registration remains valid indefinitely as long as the annual maintenance fee is paid and all regulatory requirements are met. | The Australian Register of Therapeutic Goods (ARTG) listing does not have a formal expiry date. The registration remains valid indefinitely as long as the annual maintenance fee is paid and all regulatory requirements are met. | The Australian Register of Therapeutic Goods (ARTG) listing does not have a formal expiry date. The registration remains valid indefinitely as long as the annual maintenance fee is paid and all regulatory requirements are met, meaning it remains active indefinitely provided annual maintenance fees are paid. However, the approval relies on valid supporting conformity assessment evidence (such as CE certificates or MDSAP), which typically expire after 5 years. | The Australian Register of Therapeutic Goods (ARTG) listing does not have a formal expiry date. The registration remains valid indefinitely as long as the annual maintenance fee is paid and all regulatory requirements are met, meaning it remains active indefinitely provided annual maintenance fees are paid. However, the approval relies on valid supporting conformity assessment evidence (such as CE certificates or MDSAP), which typically expire after 5 years. | The Australian Register of Therapeutic Goods (ARTG) listing does not have a formal expiry date. The registration remains valid indefinitely as long as the annual maintenance fee is paid and all regulatory requirements are met, meaning it remains active indefinitely provided annual maintenance fees are paid. However, the approval relies on valid supporting conformity assessment evidence (such as CE certificates or MDSAP), which typically expire after 5 years. |
| Registration renewal should be started this far in advance | Start the administrative process (ensuring compliance and payment) at least 1 month before June 30 to avoid issues. | It is recommended to begin the renewal preparation process a month or more in advance of the July 1st deadline | Renewal of a Class 2a medical device in Australia should be started at least 3 to 6 months before the 30 June expiry date (i.e., by March or earlier). | The annual renewal (retention) of the ARTG listing must be completed by 1 July each year to avoid cancellation. Sponsors should begin the renewal process, including paying fees and submitting declarations, at least 4-6 weeks in advance | Renewal processes like certificate updates or audits require starting preparations at least 3 to 6 months in advance |
| Complexity of the registration process for this classification | Low | Low | Low-Moderate | Moderate-High | High |
| Overall cost of gaining regulatory approval | Low | Low | Low-Moderate | Moderate | High |
Determine device classification
Classify your device according to the risk-based rules in Schedule 2 of the Therapeutic Goods (Medical Devices) Regulations 2002. Devices are classified as Class I, Is (Sterile), Im (Measuring), IIa, IIb, or III based on intended purpose and risk.
Step 1 (all classes) – Appoint an Australian Sponsor
If the manufacturer is not based in Australia, a local Australian Sponsor must be appointed.
The sponsor:
- Submits applications to TGA
- Acts as the regulatory liaison
- Must be listed on device labelling
Step 2 (all classes) – Prepare Technical Documentation
- Technical File or Design Dossier
- Evidence of conformity to Essential Principles
- Australian Declaration of Conformity
Step 3 (all classes, varies by pathway) – Provide Manufacturer Evidence
Sponsor submits Manufacturer’s Evidence (e.g., CE certificate or other conformity assessment evidence) via TGA systems for review and acceptance.
Step 4 (all classes) – Submit Medical Device Application (ARTG)
Sponsor submits an application to include the device in the Australian Register of Therapeutic Goods (ARTG). Application includes:
- Intended purpose
- Classification
- GMDN code
- Manufacturer evidence
- QMS certificates (Class IIb-Class III)
- Clinical Evaluation Report (CER) (Class IIb-Class III)
- Risk Management File (Class IIb-Class III)
- Applicable fees
Step 5 (Class IIa–III primarily) – TGA Conformity Assessment / Application Audit
TGA may conduct an application audit depending on:
- Device class
- Risk profile
- Type of evidence submitted
- Conformity assessment can vary (abridged path or full assessment)
Step 6 (all classes) – TGA Decision
- If approved → Device is included in ARTG
- If rejected → deficiencies must be addressed
Sponsor receives ARTG inclusion number
Final Step – Market Authorization & Ongoing Compliance
Manufacturer must maintain:
-
- Ongoing compliance with Essential Principles
- Valid conformity assessment evidence
- Post-market obligations